Veolia水务技术与方案

Sievers Eclipse细菌内毒素检测仪(BET)

突破性的内毒素检测技术,可实现自动化和合规性

The innovative Sievers Eclipse automated endotoxin detection platform decreases assay setup time by up to 85% and reduces Limulus Amebocyte Lysate (LAL) reagent use by up to 90% while meeting all requirements of the harmonized pharmacopeia: USP <85>, EP 2.6.14 and JP 4.01, as well as ChP 1143.

Through groundbreaking technology, the Eclipse endotoxin testing solution significantly decreases pipetting steps, reduces operator-to-operator variability, and simplifies BET test setup. The Eclipse platform leverages precise microfluidic liquid handling and embedded endotoxin to automate kinetic chromogenic assays. Throughput of 21 samples per plate is maintained without the complexity of robotics or the time and technique demands of a traditional LAL test.

下载宣传册(PDF)

合规、一致、清晰的细菌内毒素检测设备和服务

Veolia有能力让复杂的测量变得简单。

一直以来,内毒素检测需要训练有素的技术人员为凝胶凝块和基于96孔板的检测仔细准备样品和标准品。这些分析方法的准备速度很慢,容易出错和重新测试,而且不符合最新的数据完整性准则。这些方法受益于鲎试剂(LAL)检测内毒素的敏感性和特异性,但仍希望减少LAL试剂的使用。这是因为LAL是用马蹄蟹的血液制成的,人们希望尽可能地保护和优化利用这一资源。

现在,有一种更简单的方法来实现合规的内毒素检测,并通过使用微流体的新药典内毒素检测来提高可持续性。这个方法简化了测试设置,降低了重复测试率,完全符合最新的数据完整性准则,并将鲎试剂的使用量减少了多达90%。

Sievers Eclipse内毒素检测系统

特点与优势

尽管需要更简单、自动化和更可持续的内毒素检测解决方案,但许多平台无法满足关键要求或简化安装和验证以方便采用。使用Sievers Eclipse,您可以获得更简单的内毒素检测设置和完全合规的优势,无需复杂的机器人或困难的实施。

Sievers Eclipse内毒素检测平台和企业软件提供:

  • 21样品化验的设置时间为9分钟(最多节省85%)
  • LAL试剂的使用量减少多达90%
  • Compliance with all requirements of the harmonized pharmacopeia: USP <85>, EP 2.6.14 and JP 4.01, as well as ChP 1143
  • 高度定制的软件,完全符合21 CFR第11部分和ALCOA+

 

检测要求 Sievers Eclipse细菌内毒素检测仪
Requirements of USP <85>, EP 2.6.14, JP 4.01, and ChP 1143  ✔️
样品和PPC至少两份 ✔️
使用标准化内毒素(RSE/CSE)绘制至少两份的最少3点标准曲线 ✔️
阴性对照至少两份 ✔️
分析和鲎试剂批次确效至少三份 ✔️
使用FDA许可的鲎试剂 ✔️
符合21 CFR第11部分和数据完整性准则 ✔️

 

寻找长期支持?以下是Sievers Eclipse团队支持实施过程的方法:

  • IQ/OQ/PQ验证支持包(VSP)可全面验证系统,安装和验证可在数天内完成
  • 我们的团队执行完整的平台验证报告
  • 我们特定的软件协议有助于产品验证
  • 我们帮助您转换方法(例如:动态浊度法→动态显色法)
  • 我们提供桥接研究方案,帮助您从当前的动态显色平台过渡到Eclipse平台
  • 我们的可比性方案对Sievers自动化平台和96孔板平台进行了比较研究

 

系统规格

系统规格


细菌内毒素检测方法

动力学显色

检测模式

吸光度

范围,ppm

0.005–50 EU/mL

精确度

≤ 15% CV起效时间

准确度

真实值的50-200%

样品类型

水溶液,用移液管注入

校准

最长12个月

分析时间

最多2小时

样品温度

37 ± 1 °C

环境温度

17–30 °C

容量

标配:单样品流;可选:双样品流

温度控制

37 ± 0.5 °C

光源

发光二极管发射器

流体故障检测

1450 nm发射器

光学精度

≤ 5%偏离期望值

光学线性度

R ≥ 0.980

光学波长滤光片

405 nm

读取间隔

5秒

分析仪规格


输出

USB数据输出

显示

OLED

电力要求

100–240 Volts AC @50/60 Hz

保险丝

T8 A 250 VAC保险丝,尺寸为5 x20毫米。仅限Littlfuse218008或Cooper Bussmann S 506-8-R。

尺寸

高: 17.5 cm(6.9 in),宽: 35.1 cm(13.8 in),深: 50.3 cm(19.8 in)

重量

10 kg (22.1 lbs)

安全认证

UL 61010-1:2012Ed.3+R:20Apr2016             CSA C22.2#61010-2-020

CSA C22.2#61010-1-12:2012Ed.3+U1;U2    IEC 61010-1:2010 Ed.3+C1;C2

UL 61010-2-010:2015Ed.3                             IEC 61010-2-010:2003Ed.2

UL 61010-2-020:2016Ed.3                             IEC 61010-2-020:2016Ed.3

CSA C22.2#61010-2-010:2015Ed.3


最大相对湿度

85%,无冷凝

最大海拔高度

3,000 m(9,800 ft)

污染等级
2

 

选件和附件

  • 分析仪:吸光度分析仪,37°C稳定孵育控制,离心技术和安全的数据传输。

  • 微孔板:精确的液体处理设备,通过具有嵌入式内毒素标准品和阳性产品对照 (PPC) 的创新性微流技术来实现自动化。

  • 软件:高度可定制的企业解决方案,符合21 CFR PART 11和ALCOA+数据可靠性。

Eclipse微孔板与Eclipse分析仪和软件配合使用,利用离心力和气动腔在104个光学池中进行测量,并均匀分配精确量的试剂水、样品和鲎试剂。光学池包括标准曲线、样品和阳性产品对照 (PPC) 分段。将反应控制在37 +/-1C,并通过每五秒获得405 nm的读数来获得每个池的高清视图。

资源

Testimonials

“The Sievers Eclipse platform was chosen and has been an indispensable tool in our laboratory. It's easy to use design and reliability has significantly enhanced the efficiency and accuracy of our large-scale water validation projects. The instrument's software interface is intuitive, and the technical support provided by Sievers has been exemplary, ensuring smooth operation whenever needed. The Sievers Eclipse has been the game changer in our laboratory.”

Dr. Cheryl Andreoli
Lab Director
Technical Safety Services, LLC

 

“A standout feature of the Sievers Eclipse BET Platform is the intuitive software. The platform’s software with step-by-step prompts is remarkably straightforward and easy to follow. The ease of use means our team can focus more on interpreting results and less on navigating complex software. In addition the software integrated seamlessly with our LIMS system for an entirely paperless endotoxin testing process from sampling to results entry and review.”

Lindsay Matern
QC Microbiology Lab Manager
Selkirk Pharma

 Schedule a demo

什么是细菌内毒素检测?

细菌内毒素检测是制药、医疗器械和生物技术行业的一项重要质量控制程序,用于检测和量化各种产品中是否存在内毒素。内毒素是革兰氏阴性细菌细胞外膜的组成部分,即使含量很少,也会令人类和动物发生严重的免疫反应。

什么是细菌内毒素?

细菌内毒素存在于革兰氏阴性菌的细胞壁外膜中。革兰氏阴性菌的内毒素作用主要是由于脂多糖(LPS)的脂质A成分。当在血液和其他体液中超过一定浓度时,脂质A成分可以产生热原(引起发烧)反应,这可能是危险甚至致命的。由于某些药品和医疗设备会进入血液,因此这些行业需要进行内毒素检测。由于具有强大的毒性而且会对患者安全造成影响,内毒素受到了美国FDA和国际药典的全球监管,适用于任何与人类和/或动物血液接触的生命科学产品。

为什么要检测水或最终产品中的细菌内毒素?

在制药、医疗设备和其他生命科学行业,内毒素检测对于患者安全和质量控制至关重要。内毒素测试是在制药级水系统上进行的,包括用作配料水的注射用水(WFI),以及药品、动物药品、生物和医疗器械产品等成品。测试的药品和医疗产品包括肠外药物产品(通过注射给药,如静脉注射、皮下注射、肌肉注射和皮内注射)以及与血液、心血管系统、淋巴系统或脑脊液直接或间接接触的医疗器械。全球药典(USP <85>、EP 2.6.14和JP 4.01)描述了对细菌内毒素检测(BET)和内毒素限量的要求。FDA指南为生物制品、药品和器械制造商提供了有关药典程序中涵盖的测试建议以及有关验收标准的当前思路建议。

细菌内毒素测试方法是如何工作的?

美国FDA接受鲎试剂(LAL)测试作为内毒素的成品测试已有近40年历史。这是一种替代兔热原检测的体外检测方法,可检测细菌内毒素的存在和浓度。在鲎试剂测试中,马蹄蟹(Limulus Polyphemus)血液中的裂解物会与细菌内毒素发生反应,以表明样品中存在内毒素。LAL检测有三种细菌内毒素检测原理方法:凝胶凝块法、比浊法和显色法。在凝胶凝块法中,操作员将LAL和样品混合,然后筛选反应管中的凝胶形成。由于测试结果是来自对凝块形成的主观解释,因此凝胶凝块法是一种定性测试。在比浊法和显色法中,操作员将样品与LAL试剂混合,并测量颜色变化或浊度随时间的变化。显色和比浊鲎试剂测定是定量的,可显示存在的内毒素量,测试结果是根据标准曲线计算得出的。全球药典、USP <85>、EP 2.6.14和JP 4.01中描述了进行BET的内毒素限量和要求。《欧洲药典》中一个新的通用章节介绍了一种使用重组因子C (rFC) 代替经典LAL方法的BET方法。

Sievers Eclipse BET平台需要使用多少鲎试剂?

只需使用1 mL LAL鲎试剂,即可在Eclipse平台上运行21个样品。通过减少高达90%的马蹄蟹(HSC)裂解液,Eclipse减少了对这种宝贵自然资源的需求,并提供了可以维持全球HSC种群的完全合规的BET检测。

Sievers Eclipse BET平台是否兼容?

是的。Eclipse平台使用经FDA许可的市售LAL,符合全球统一药典、USP <85>、EP 2.6.14和JP 4.01的所有要求。使用Eclipse平台进行的细菌内毒素检测包括:使用标准化内毒素的最小三点标准曲线(一式两份);样品和PPC(一式两份);阴性对照(一式两份);分析仪和裂解液批次鉴定(一式三份);使用经FDA许可的LAL。它还符合21 CFR第11部分和数据完整性准则。

​​​​​​​标准曲线是否可以在Sievers Eclipse中自动实现?

是的。传统上,对于符合要求的内毒素检测,最终用户必须从标准内毒素库存瓶中构建至少一式两份三点标准曲线;必须有重复的阴性控制;每个样品和PPC必须一式两份。有了Sievers Eclipse BET平台,这些步骤可以通过使用预嵌入的内毒素标准品实现自动化,这些内毒素标准品包括5点标准曲线和预嵌入的PPC。因此,用户所需要做的就是将细菌内毒素检测用水和样品置入平板,而不需要额外的准备工作。其结果是实验设置仅需9分钟,而其他平台需要60分钟以上。

 索取信息与询价